Fluorescence & Phosphorescence

Drag to turn the view. Scroll or pinch to zoom. Double-click to reset.

Key moments

When you press play, the animation slows down and then freezes completely for 6 to 8 seconds at the moments that matter: the flash, the first molecule that loses its energy as heat, the first photon leaving, the first crossing to the triplet, the first collision with a quencher, the jump up and the drop back down in the bond picture, and the turning points in the cuvette and phosphorescence tabs. Every moment is taken from the tab’s own simulation, so the molecule ringed in red is one that really did that, and the moments move when you change the settings. The molecule and its caption pulse in red, everything else fades, and the close-up shows the same molecule. A red bar under the caption shows how long until the animation moves on. Each held event is one of many happening all the time. In the first and third tabs the clock starts 3 seconds before the flash, so you can see the molecules unexcited first. Red diamonds on the timeline mark the moments; “Next ▸” jumps to the next one, and “Key moments” turns the pauses off. Dragging the time slider or playing at 2× skips them. The Practice tab has no moments.

Where the energy goes

A short flash excites 200 molecules at once. Press play and the clock runs to five lifetimes (slowed about a billion times). Each molecule leaves the excited state S₁ at a random moment by exactly one path: it emits a photon (fluorescence, rate constant kf), turns the energy into heat in the solvent (internal conversion, kic), or flips a spin and crosses to the triplet (intersystem crossing, kisc). The paths compete, so the quantum yield is Φ = kf/(kf + kic + kisc) and the lifetime is τ = 1/Σk. The close-up follows one molecule’s electron. The Jablonski diagram draws each arrow with a width proportional to its rate. The graph counts photons as they arrive and compares them with the decay expected from the rates; the counts scatter around it because each molecule’s fate is random. kf = Φ/τ is computed from cited Φ and τ. The starting kisc uses the measured triplet yield where known: anthracene loses most of its energy to the triplet (ΦT ≈ 0.71, ΦF = 0.27), fluorescein about 2%. For quinine, rhodamine 6G and tryptophan the non-radiative loss is not split and is drawn as heat. Tryptophan’s decay is really two-exponential (about 3.1 and 0.5 ns); 2.8 ns is used as an average. Idealizations: one excitation per molecule, no quenchers, vibrational levels drawn for shape only.

Why the light is redder

Absorption is so fast that the nuclei do not move (vertical arrow, Franck–Condon principle), so the molecule lands high on the S₁ curve. Within picoseconds it slides to the bottom, giving heat to the solvent; only then, nanoseconds later, does it emit, landing high on the S₀ curve. Both steps take energy away from the photon, so emission is at longer wavelength: the Stokes shift, Δν̃ = 10⁷/λabs − 10⁷/λem (6349 cm⁻¹ for quinine). Because emission always starts from the lowest excited state (Kasha’s rule), moving the excitation wavelength changes how bright the emission is but not where it is (Kasha–Vavilov rule). Anthracene shows the mirror image between its absorption and emission bands. Band positions are cited; band shapes, the vibronic spacing (1400 cm⁻¹) and the solvent shifts are illustrative. The potential curves are schematic.

Quenching

Dynamic quenching: a quencher (I⁻ or O₂) collides with an excited molecule during its lifetime and removes the energy without light. Intensity and lifetime fall together: F₀/F = τ₀/τ = 1 + KSV[Q], with KSV = kqτ₀. For fluorescein and iodide KSV = 9.6 L/mol, so 0.10 M I⁻ gives F₀/F = 1.96 and τ = 2.04 ns, and kq = 2.4 × 10⁹ L mol⁻¹ s⁻¹, close to the diffusion limit. Static quenching: a dark ground-state pair forms before excitation; those molecules never emit, but the free ones keep τ₀, so τ₀/τ stays 1. With both, F₀/F = (1 + KD[Q])(1 + KS[Q]) curves upward. The static constant (9.6 L/mol) is illustrative, chosen equal to the dynamic one for comparison. The quencher is drawn in purple; molecules are not to scale.

Inside the cuvette

In a standard fluorimeter the detector looks at the middle of a 1 cm cell at 90° to the excitation beam. The beam weakens as 10−εcx on its way in, so the glow is strongest at the front face. Dilute samples give a signal proportional to concentration; as concentration rises the front of the cell absorbs the light before it reaches the viewed window (the primary inner-filter effect). For quinine (ε ≈ 5700 L mol⁻¹ cm⁻¹ at 350 nm) viewed from 0.4 to 0.6 cm, the signal is 5% below the straight line already at an excitation absorbance of 0.045, peaks at 154.5 µM (A = 0.88) and then falls, so one signal can match two concentrations. Keep A below about 0.05 or dilute. The common correction F × 10A/2 works only below A ≈ 0.3. Front-face viewing levels off instead of falling. Emission reabsorption (secondary inner filter) is zero here because quinine does not absorb at 450 nm.

Phosphorescence

From S₁ a molecule can flip one electron spin and cross to the triplet T₁. Returning to S₀ needs another spin flip, a forbidden step, so the triplet lives milliseconds to seconds and emits at longer wavelength than the fluorescence. Heavy atoms (iodine, bromine) speed up spin flips: erythrosin B, fluorescein with four iodines, loses almost all its fluorescence to the triplet. O₂, itself a triplet, quenches long-lived states very efficiently: with kq ≈ 10¹⁰ L mol⁻¹ s⁻¹ and 0.26 mM O₂ (air-saturated water), a 5 ms triplet is shortened about 13 000-fold while a 4 ns fluorescence changes by 1%. That is why phosphorescence is measured in rigid glasses at 77 K or in degassed solutions. A delay after the flash (a rotating chopper or a gated detector) removes the fluorescence completely: a 0.1 ms delay keeps e−0.1/5 = 98.0% of a 5 ms phosphorescence. Benzophenone and erythrosin values are approximate; the erythrosin phosphorescence yield (0.01) is illustrative; real room-temperature phosphorescence yields in fluid solution are very small. Erythrosin’s fluorescence lifetime in water is 89 ps.

Measure samples

A spectrofluorimeter: xenon lamp, excitation monochromator, cuvette, emission monochromator at 90° and photomultiplier. An emission scan holds the excitation wavelength and steps the emission; an excitation scan holds the emission and steps the excitation, and its shape follows the absorption spectrum. Wider slits give more light but broader bands. Six samples: quinine sulfate in 0.1 M HClO₄, fluorescein in 0.1 M NaOH (Φ 0.925), rhodamine 6G (Φ 0.95) and anthracene (Φ 0.27) in ethanol, tryptophan in water and benzophenone (no fluorescence at room temperature). The calibration panel reads five standards (0.2–1.0 µM) and one unknown at the current wavelengths. Φ, λ and ε are published values (approximate where marked in the research notes); band shapes and the count scale are illustrative, and noise is not simulated.

Practice

Ten kinds of question rotate: quantum yield from rates, kf from Φ and τ, Stokes shift, Stern–Volmer, the quenching rate constant, dynamic or static, the inner-filter maximum, gated phosphorescence, oxygen quenching and an unknown from a standard. Every number and equation you need is in the Conditions box. A wrong answer is crossed out and you can try again; Hint gives the method; asking for the solution needs two confirmations. ★ counts answers right on the first try. The 3D view shows the tab each question comes from; press play to run it.

Sources

Φ(quinine, 0.1 M HClO₄) = 0.60 ± 0.02: Brouwer, Pure Appl. Chem. 83, 2213 (2011); Nawara and Waluk, Anal. Chem. (2019). Φ(fluorescein, 0.1 M NaOH) = 0.925 ± 0.015 and Φ(rhodamine 6G, ethanol) = 0.950 ± 0.015: Magde, Wong and Seybold, Photochem. Photobiol. 75, 327 (2002). Anthracene ΦF = 0.27, ΦT ≈ 0.71; tryptophan Φ ≈ 0.13, ε(280) ≈ 5600; erythrosin B τ = 0.089 ns (lifetime-standard compilation); benzophenone phosphorescence 5–6 ms at 77 K. Fluorescein–iodide KSV = 9.62 ± 0.27 L/mol: J. Chem. Educ. (2015), open-box iPad fluorescence quenching. Lifetimes, kq for O₂ and the inner-filter correction: Lakowicz, Principles of Fluorescence Spectroscopy, 3rd ed.; Tucker, Amszi and Acree, J. Chem. Educ. (1992). ε(fluorescein) = 76 900: Sjöback et al. (1995). Other ε and λ: PhotochemCAD (OMLC). Benzophenone phosphorescence: Turro, Modern Molecular Photochemistry. Skoog, Holler and Crouch, Principles of Instrumental Analysis, Ch. 15; Harris, Quantitative Chemical Analysis, Ch. 18. Air-saturated O₂ in water at 25 °C: 0.26 mmol/L. Slowed time, schematic curves and molecule sizes are labelled where they appear.