HexaMoleInteractive Chemistry

Light & Absorbance Lab

Drag to turn the view. Scroll or pinch to zoom. Double-click to reset.

Key moments

In two tabs the animation slows down and then freezes for 6 to 8 seconds at a turning point: in the Spectrophotometer when the scan reaches the sample’s λmax, where it absorbs most, and in Where Beer’s law bends when the reading falls 1% and then 10% below the straight line. Both are computed from the tab’s own measurement model without the random noise, so the reading on screen scatters a little around the value in the caption, and the moments move when you change the sample, slit, stray light or dimerization. The caption pulses in red with a countdown bar. Red diamonds on the timeline mark the moments; “Next ▸” jumps to the next one, and “Key moments” turns the pauses off. Dragging the time slider or playing at 2× skips them. The photon tab has no moments: its photons are absorbed at random as you watch.

Light

Light is a wave and a stream of photons. Wavelength λ and frequency ν are linked by c = λν, and each photon carries E = hν = hc/λ: shorter wavelengths carry more energy. Amplitude sets the brightness (intensity ∝ amplitude²), which means more photons, not more energetic ones. Drag the marker on the spectrum strip, or use the slider, to move from the ultraviolet through the visible to the near infrared.

Spectrophotometer

White light from the lamp is spread by a grating in the monochromator, and the exit slit picks a narrow band of wavelengths. That band passes through the cuvette and reaches the detector. First measure a blank (solvent only) to set 100% transmittance; then the sample gives T = I/I₀ and A = −log T. Press play to scan the whole spectrum point by point, as a scanning instrument does, and use “Go to λmax” to find the strongest absorption. The slit width sets how many wavelengths pass at once.

Photons and Beer’s law

Photons cross the cuvette, and in each thin slice every photon has the same small chance of being absorbed. So each slice removes the same fraction of the light reaching it: the intensity falls exponentially through the sample, and the absorbance, its logarithm, rises in a straight line with depth. That is Beer’s law, A = εbc. The dots on the graphs are the photons actually counted.

Colour and spectra

A solution’s colour is the light it lets through. The lab computes it from the absorption spectrum, so a sample absorbing green looks purple, the colour opposite on the colour wheel. Raising the concentration deepens the colour but keeps the hue. For bromothymol blue, the acid and base forms absorb differently; their spectra cross at the isosbestic point, where absorbance does not depend on pH.

Where Beer’s law bends

Real measurements leave the straight line. Stray light reaching the detector caps the reading near −log s. A slit wider than the absorption peak averages over weaker wavelengths and reads low. Methylene blue pairs up into dimers at high concentration, which absorb less at 664 nm. Detector noise gives the smallest relative error near A = 0.434, which is why measurements are kept between about 0.1 and 1.

Calibration and mixtures

Six standards are made by diluting a stock and measured at λmax with realistic noise; the least-squares line gives the slope (εb), the intercept, R², and the limits of detection (3sy/m) and quantification (10sy/m). An unknown read three times gives a concentration with its uncertainty. In the mixture experiment, permanganate and dichromate both absorb at 440 and 545 nm; because absorbances add, two wavelengths give two equations to solve for both concentrations.

Phone colorimeter

Turn a phone or laptop camera into a colorimeter. Place the sample (a cuvette or a clear vial) in front of a white card, lit evenly, and drag the red box onto the sample and the green box onto the card. Measure a blank, then standards of known concentration, then your unknown. To make the camera trustworthy, the lab uses the rear camera and tries to lock exposure, white balance and focus; converts the camera’s gamma-encoded pixels back to linear light; divides the sample by the white card in every frame, so changes in lighting or exposure cancel; averages many frames, giving each reading an uncertainty; and ignores saturated pixels. It picks the colour channel the sample absorbs most. A camera channel spans a wide band of wavelengths, so keep absorbance below about 1 for a straight line. The camera works on the published page in a current browser; nothing is recorded or sent anywhere.

Practice

Forty kinds of question on light, Beer’s law, the instrument, colour, and calibration, with new numbers each time. Wrong answers can be retried, a hint points to the method, and the full solution is there if you need it.

Settings used

Physical constants are exact SI values. Permanganate and dichromate spectra are fitted to standard data (permanganate λmax 525 nm, ε ≈ 2.4 × 10³ M⁻¹cm⁻¹; the classic mixture values at 440 and 545 nm). Copper(II), methylene blue, bromothymol blue and coffee spectra are approximate shapes from published spectra, and the methylene blue dimerisation constant is illustrative. Solution colours are computed from the transmitted spectrum with the CIE 1931 colour-matching functions, so they are close to, but not exactly, what you would see.